"Evaluation Of Safety And Performance Of WALL-NUT™ Left Atrial Appendage Occluder System (LAAOD): A Porcine Model Analysis Over Multiple Time Points"

Abstract

This study evaluated the thrombogenicity and biocompatibility of the WALL-NUT™ left atrial appendage (LAA) occluder device in three animals (porcine model). No device-related thrombogenic events, embolization, or adverse outcomes were observed throughout the study period. Transesophageal echocardiography (TEE) revealed stable device positioning without evidence of thrombus formation, device migration, or pericardial effusion. Clinical assessments showed no signs of morbidity, abnormal weight loss, or significant changes in hematology or clinical chemistry between pre and post-implantation. Histopathological evaluations indicated a mild inflammatory response around the implant, with progressive endothelialization of the device surface. Two animals exhibited up to the 75% endothelial coverage, while one achieved complete coverage by 90 days. The presence of inflammatory cells was noted, but without escalation of the response over time. These findings demonstrate the LAA occluder’s strong safety profile and biocompatibility, supporting its potential for clinical application in preventing stroke in patients with atrial fibrillation. Future studies with longer follow-up are recommended to confirm long-term efficacy.

Key Words

WALL-NUT™ Left atrial appendage (LAA) occluder Device, Porcine Model, Safety Evaluation, Performance Assessment, radiography and angiographic findings, Histopathological Findings.

Cite This Article

""Evaluation Of Safety And Performance Of WALL-NUT™ Left Atrial Appendage Occluder System (LAAOD): A Porcine Model Analysis Over Multiple Time Points"", International Journal of Emerging Technologies and Innovative Research (www.jetir.org), ISSN:2349-5162, Vol.11, Issue 12, page no.f178-f201, December-2024, Available :http://www.jetir.org/papers/JETIR2412522.pdf

ISSN


2349-5162 | Impact Factor 7.95 Calculate by Google Scholar

An International Scholarly Open Access Journal, Peer-Reviewed, Refereed Journal Impact Factor 7.95 Calculate by Google Scholar and Semantic Scholar | AI-Powered Research Tool, Multidisciplinary, Monthly, Multilanguage Journal Indexing in All Major Database & Metadata, Citation Generator

Cite This Article

""Evaluation Of Safety And Performance Of WALL-NUT™ Left Atrial Appendage Occluder System (LAAOD): A Porcine Model Analysis Over Multiple Time Points"", International Journal of Emerging Technologies and Innovative Research (www.jetir.org | UGC and issn Approved), ISSN:2349-5162, Vol.11, Issue 12, page no. ppf178-f201, December-2024, Available at : http://www.jetir.org/papers/JETIR2412522.pdf

Publication Details

Published Paper ID: JETIR2412522
Registration ID: 552816
Published In: Volume 11 | Issue 12 | Year December-2024
DOI (Digital Object Identifier):
Page No: f178-f201
Country: Vapi, Gujarat, India .
Area: Science & Technology
ISSN Number: 2349-5162
Publisher: IJ Publication


Preview This Article


Downlaod

Click here for Article Preview

Download PDF

Downloads

000220

Print This Page

Current Call For Paper

Jetir RMS